5080A High Compliance Multi-Product Calibrator

$111.00

Spesifikasi: 5080A High Compliance Multi-Product Calibrator

Technical Specifications
Voltage DC Range: 0 to ±1020V
Best 1 Year Specification: ±0.025% of setting
Current DC Range: 0 to ±20.5A
Best 1 Year Specification: ±0.025% of setting
Resistance Range: 0 to 190 MO, 19 values in x1 and x1.9
Best 1 Year Specification: ±0.025% of setting
Voltage AC Range: 1 mV to 1020V
45Hz to 1 kHz, Sine
Best 1 Year Specification: ±0.25% of setting
Current AC Range:29 mA to 20.5A
45 Hz to 1 kHz, Sine
Best 1 Year Specification: ±0.25% of setting
DC Power Range:10.9 mW to 20.5 kW
Best 1 Year Specification: ±0.125% of setting
AC Power Range: 10.9 mW to 20.5 kW
45-65 Hz, PF=1
Best 1 Year Specification: ±0.36% of setting
Phase Range: 0 to ± 179.99°
45 Hz to 1 kHz
Best 1 Year Specification: ±0.25° of setting
Frequency Range: 45 Hz to 1 kHz
Best 1 Year Specification: ±50 ppm of setting
General Specifications
Warm up Time  Twice the time since last warmed up, to a maximum of 30 minutes
Settling Time Less than 5 seconds for all functions and ranges except as noted
Standard Interfaces RS-232 and Ethernet
Temperature Performance Operating: 0°C to 50°C
Calibration (tcal):15°C to 35°C
Storage:-20°C to +70°C
Temperature Coefficient Temperature coefficient for temperatures outside tcal ±5 °C is 10 % of the stated specification per °C for temperatures in the range of 0 °C to 35 °C. Above 35 °C, the temperature coefficient is 20 % of the stated specification per °C.
Relative Humidity Operating: <80% to 30°C, <70% to 40°C, <40% to 50°C
Storage:<925%, non-condensing
Altitude Operating: 2,000m (6,500 ft) maximum
Non-operating: 12,200m (40,000 ft) maximum
Safety Meets EN. 61010-1:2001, CAN/CSA-C22.2 No. 61010-1-04, UL 61010-1: 2004
Analog Low Isolation 20V
EMC Meets EN613261:2006.
Line Power Line Voltage (selectable): 100V, 120V, 220V, 240V
Line Frequency: 47 to 63 Hz
Line Voltage Variation: + 25% about line voltage setting
Power Consumption 600 VA
Calibration Documentation Report of calibration with data included; 17025 accredited report optional
Dimensions Height:19.3 cm (7.6 in)
Width:43.2 cm (17 in), 44.3 cm (17.5 in) including handles
Depth:53.8 cm (21.2 in)
Weight (without options) 22 kg (48 lbs.)
Absolute Uncertainty Definition
The specifications include stability, temperature coefficient, linearity, line and load regulation, and the traceability of the external standards used for calibration. It is not necessary to add anything to determine the total specification for the temperature range indicated.
Specification Confidence Interval 99%
SKU: 5080A High Compliance Multi-Product Calibrator Kategori: Brand:

Cv Brema Perkasa Abadi Adalah distributor resmi Fluke di Indonesia yang menyediakan segala produk Fluke

5080A High Compliance Multi-Product Calibrator

Fitur utama

  • Calibration solutions for both analog and digital workloads with high compliance and protection circuitry.
  • Calibrate analog meters, 3.5 and 4.5 digits DMMs, panel meters, watt meters, clamp meters, oscilloscopes up to 200 MHz, and more.
  • Equipped with versatile software applications for paperless results recording.
  • Built-in protection circuitry to safeguard against damaging input voltages.